SPECIFICATIONS

Model PD1000I
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14.7 ounces; 16.3 ounces with battery
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.75"L x 3.4"W x 2.8"H (12.06 cm L x 8.64 cm W x 7.11 cm H)
Model PD1000G
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18.9 ounces; 20.5 ounces with battery
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.88"L x 3.4"W x 2.8"H (12.4 cm L x 8.64 cm W x 7.11 cm H)
All Models
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (2) Standard "AA" alkaline or NiMH batteries.
Operational Voltage Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3 to 3.6V DC
Power Requirements . . . . . . . Average steady state "ON" current 1.6 uA. Batteries other than alkaline or NiMH
Degree of Protection Against Electric Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TYPE BF applied part
Modes of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous / Pulsed
Operating Temperature Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14° to 104°F (-10° to 40°C)
Operating Pressure Range. . . . . . . . . . . . . . . . . . . . . . . . . . . .500 to 2250 PSIG (34 to 155 bar—tank pressure)
Operating Atmospheric Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 to 1020 hPa
Operating Humidity Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 95% R. H. non-condensing
Storage and Transportation Temperature Range (Tested at ~933 hPa) . . . . . . . . . .-40° to 158°F (-40° to 70°C)
Storage and Transportation Humidity Range (Tested at ~933 hPa) . . . . . . . . . . 0 to 95% R. H. non-condensing
Expected Shelf and Service Life (excluding batteries) . . . . . .5 years based on 4 hours use per day at 20 BPM
Degree of Protection Against Ingress of Liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IPX1
Safety Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . meets ISO 18779: 2005(E)
Approval Body And Safety Standard . . . . . . . Approved by CSA to: IEC 601-1; CAN/CSA-C22.2 No 601.1-M90
US Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4,519,387; 5,755,224; 4,457,303
®
Certified to CAN/CSA C22.2 No. 601.1-M90
SE-1000
are not recommended due to the capacity needed for operation and battery life of
the unit. Typical new battery life is 200 hours when used at 25°C, 2 LPM and 20
BPM. Settings and breath rate will affect battery life. After the Low Battery (flash-
ing red) light illuminates, the unit will continue to operate about four hours when
used at 25°C, 20 BPM and the 6 LPM setting. Settings, breath rate, and battery
conditions will affect use times. Refer to local regulations for battery recycling and/
or disposal requirements.