Determining the required cleaning and disinfecting level
WARNING
The level of cleaning and disinfecting required for the system is dictated by the type of tissue it contacts
during use. Use
Table 1
Table 1: Choosing a cleaning and disinfecting method
Did any part of the system or transducer come in contact with broken skin, blood, mucosal
membranes, or bodily fluids?
Came in contact with broken
YES
skin, blood, mucosal membranes,
or bodily fluids.
OR
Did not come in contact with
NO
broken skin, blood, or bodily
fluids.
Spaulding classifications
Spaulding classifications (non-critical, semi-critical) determine the approach for cleaning and disinfecting
medical equipment based on the device, the way it has been used, and the risk of infection. The system and
transducers are designed for use within the Spaulding classifications of non-critical and semi-critical uses.
Cleaning and disinfecting
The cleaning instructions contained in this section are based on requirements
mandated by the American Food and Drug Administration (FDA). Failure to
follow these instructions may result in cross contamination and patient infection.
to determine the level of cleaning and disinfecting required.
Option A
"Clean and disinfect system and
Go to
transducer to a high level (semi-critical
uses)"
on page 10.
Option
B
"Clean and disinfect system and
Go to
transducer to a low level (non-critical
uses)"
on page 15.