Measuring Temperature
With probes (adapter and probes are optional)
Connect the thermocouple adapter to the two input terminals of the clamping ; make sure polarity
is respected.
(- on the COM terminal and + on the + terminal)
Connect the desired probe to the adapter.
712-01
Adapter
DX.12.06
Flexible probe
DX.12.08
Bracelet probe
DX.12R
Extension (is used with the bracelet probe)
DX.12.11
Standard needle probe
DX.12.15
Surface contact probe
DX.12.17
Air probe
The T° EXT symbols appear on the display.
DX.12-11
DX.12-15
Internal Temperature
Probe Type
Integrated Circuit
Display Range
° C
Measurement
-10° C to + 50° C
Range
+15° F to +120° F
Precision
± 1,5° C
± 2,7° F
Resolution
0,1° C
0,1° F
20
DX.12-06
DX.12-08
DX.12-11
DX.12R
DX.12-15
DX.12-17
712-01
DX.12-06
DX.12-08
DX.12-17
External Temperature External Temperature
K Couple
K Couple
° C
° C
-50.0° C to + 399,9° C
+400° C to + 1,000° C
-50° F to + 399,9° F
+400° F to + 1,800° F
1% L ± 1,5° C
1% L ± 1,5° C
1% L ± 2,7° F
1% L ± 2,7° F
0,1° C
1° C
0,1° F
1° F