Calibration of temperature sensor in a new probe
for HMM
Attention: Doesn't concern a new device or a de-
vice sent from repair service.
Procedure:
1. Before starting calibration process it is necessary to
prepare:
– a bucket/bowl etc. filled with water and ice. Ice
flowing in the water guarantees that the water
will have 0 degrees Celsius.
– a bucket/bowl etc. filled with warm water
– a thermometer
2. Attach the new probe to the main unit. Put the
probe into the bucket with warm water and leave it
inside for about 5 minutes until the probe will reach
temperature similar to the temperature of water.
Put the thermometer into the bucket and measure
the temperature of water.
3. Press and hold 4 buttons: „MAX", „MIN", „T/M"
and „ON/OFF". Keep holding the buttons for about
1 second. Release all buttons, letters „CAL" will be
displayed.
18

CHAPTER 5 | CALIBRATION Of TEMPERATURE SENSOR

4. Using „MAX" and „MIN" buttons set the number
233 and press „T/M".
5. Using „MAX" and „MIN" buttons choose „HI" and
press „T/M".
6. If the temperature displayed on the screen is dif-
ferent from readings of the thermometer, correct
it and set proper temperature with „MAX", „MIN"
buttons and confrm it with „T/M".
7. Put the probe into the bucket filled with water and
ice and leave it inside for about 5 minutes until the
probe will reach temperature similar to the tem-
perature of water.
8. Using „MAX" and „MIN" buttons choose „LO" and
press „T/M".
9. If displayed temperature is different from „0", mod-
ify it with MAX and MIN buttons, until you reach
„0". Confirm with T/M.
10. Turn the device off. The temperature sensor in the
new probe is now properly calibrated.