Modifying the moisture indications readings
This option is used for the modification (adjustment)
of moisture readings by increasing or decreasing the
displayed value (of the device) by the same value in
the entire measuring range. However, by modifying
the displayed values we are changing the measuring
range limits, i.e., when we make an adjustment by
„+2%" the measuring range will change from 10%-
80% to 12%-82%.
The modification is made, if the user finds that for a
given force of the pressed bale the tool has a ten-
dency to over- or understate the results by the similar
value in the entire moisture range. This modification
is useful when the bales of hay/straw are poorly or
very hard pressed, because the HMM curve is based
on an average compression strength.
Procedure:
1. Press and hold four buttons at the same time:
"MAX", "MIN", "T/M" and "ON/OFF".
12

CHAPTER 3 | MODIfYINg THE MOISTURE INDICATIONS READINgS

2. Hold 4 buttons pressed for about 1 second.
Then release the buttons – letters "CAL" will be
displayed.
3. Using "MAX" and "MIN" buttons set the number
399 and press "T/M".
4. When the display shows „0" the device is ready for
calibration.
5. Enter the requested adjustment, keeping the
range from – 5 % to +5 % (e.g. „+2" or „–5")
6. Confirm the entered value with „T/M" button,
then turn off the device.
An example of the correctly entered modification
1. When you have entered „–5", then each time the
device is turned on, the display will show the fol-
lowing message „CAL"–5. This means that the de-
vice is changing the measuring range from
10%-80% to 5%-75%.