2. Setting of Air & Water Supply Pressure
Measure the supply pressure at the handpiece / hose connection point and set the pressure to the value
specified on the specification table. ( Fig. 1 )
*Multi Gauge : Midwest 4 Hole Order code Z109400
Borden2/3 Hole Order code Z109200
WARNING
Do not exceed the optimum pressure specified on the specification table.
CAUTION
Do not use air contaminated by dust, moisture and oil.
4
Exhaust
Chip Air Pressure
Air Pressure
Water Pressure
Drive Air
Pressure
Handpiece
Fig. 1