CAUTION
· Make sure that the O-Ring is removed, when removing the Cartridge. O-Ring might be remained in the Head, when removing the Cartridge. If
there is a remained O-Ring, the new Cartridge can not be inserted properly.
· Use only NSK genuine cartridge. If you use other cartridge, specification may not be satisfied and we can not warranty of this situation.
· The head cap screw thread is very fine. To prevent damage to the thread always first tighten the handpiece head cap using fingers, then
secure with the head cap wrench.
· Make sure that the O-ring (Fig. 12) is correctly located on the cartridge.
Worn or damaged O-ring may cause air or water to leak.
When water is found in the exhaust air or water leak from the coupled point, replace our O-rings as shown in
Fig.13.
*Replacement O-ring Set (4pcs) : Order Code Y900592
O-Ring
Fig. 13
15