set up
the dot on the receptacle. To disconnect: grasp the plug, twist counterclockwise, and pull.
4.
Ensure the DC accessory power port is clean and a good connection can be made. Insert the other end of the DC Adapter into the vehicle's
DC accessory power port. When properly connected and receiving power from the DC source, a green indicator light on the DC Adapter
and the External Power Present light on the control panel will illuminate.
5.
Secure the iGo and the DC Adapter in your vehicle and make sure the air inlet and exhaust vents are not blocked.
NOTE–
The battery will not charge when connected to a DC accessory power port.

Battery Operation

Electronics are static sensitive and may be an electric shock hazard� When the battery is removed, do not touch the contacts
in the Battery Bay or on the battery� The communication port is also static sensitive and should not be touched�
The iGo can also be powered by a Rechargeable Battery. If the battery is installed and adequate power is available, the iGo will charge the battery
any time AC power is present and the battery temperature is below a safe charging temperature. If external power is disconnected, the iGo will
automatically switch over to the battery (if installed). When external AC power is restored, the iGo will accept power from the external source and
charge the battery.
Installing the Battery: Insert the battery tabs into the slots in the Battery Bay and push the battery until it is secure and flush with the back of the
system. When the battery is properly installed and the iGo is On, the Battery Status Gauge will appear on the Control
Panel. (If you DO NOT have a Battery installed, the Status Gauge will not light up).
Removing the Battery: Push down on the Battery latch. The Battery will tilt so that you can grasp it for removal.
CAUTION
Battery Capacity: The capacity of the Battery is lit up on the control panel. A variety of factors, such as age of battery, flow rate, and PulseDose
or Continuous Flow Mode operation, determine the duration of operating time. The following table provides estimates of
amount of time that a new, fully charged iGo Battery will operate.
Typical New Battery Operation Time
Setting
Continuous Flow
1.0
4.0 hours
2.0
2.4 hours
3.0
1.6 hours
4.0
5.0
6.0
NOTE
–Be sure to check the battery charge level before travel; battery will discharge over time.
Do not short circuit the Battery's metal contacts with metallic objects such as keys or coins� It may cause sparks or excessive
heat�
Do not tamper with, disassemble, puncture, or crush the Battery� There are no serviceable parts inside the Battery� Do not
open� Battery electrolytes may be toxic if swallowed and can be harmful to skin and eyes� Keep the Battery away from
children�
Exposing the Battery to water or other liquids may cause personal injury�
Replace the Battery with only approved DeVilbiss Rechargeable Battery� The iGo system can only work with a DeVilbiss
Battery� Use of a damaged battery or a non-approved Battery may damage the unit, present a risk of fire or explosion, cause
personal injury, and will void the warranty�
The Battery used in this device may present a risk of fire or chemical burn if mistreated� DO NOT disassemble, incinerate, or
heat above 140°F (60°C) such as in a vehicle parked in the sun or on a hot day�
Return your Battery to your provider for proper disposal�
CAUTION–DO NOT drop the Battery or expose it to mechanical shock.
CAUTION–Only use the Battery for its intended purpose.
NOTE–
Operating time will degrade with Battery use and age.
NOTE
– Store your Battery in a cool, dry location when not in use. This will help to assure the longevity of your Battery.
NOTE
– Charging the Battery below 50°F (10°C) or above 104°F (40°C) may degrade the performance. The Battery will not charge above 45°C.
� WArning
– W hen the battery is removed, the Communication Port is exposed. Do not attach any cables to this port;
it is for service only. The Battery Contacts inside the bay and on the battery are also exposed and should
not be touched.
PulseDose (20 BPM)
5.4 hours
4.7 hours
4.0 hours
3.5 hours
3.2 hours
3.0 hours
� WArning
EN
A-306
9