Agilent Technologies TwisTorr Medium-TMP Manual De Instrucciones página 151

Ocultar thumbs Ver también para TwisTorr Medium-TMP:
Fault: this open collector output signal is ON when a system fault
condition is detected.
Programmable set point: this open collector output signal is enabled
when the reference quantity chosen (frequency, current or time) is
higher than the set threshold. The signal can be "high level active" (that
is the output is normally at 0 Vdc and becomes 24 Vdc when activated),
or "low level active" (that is the output is normally at 24 Vdc and
becomes 0 Vdc when activated). Moreover, if the reference quantity is
the frequency or the current drawn, it is possible to set the hysteresis
(in % of the threshold value) to avoid bouncing.
For example:
Figure 8
The set point output stays at 0 Vdc until the frequency becomes
higher than 505 Hz (that is 500 Hz + 1% of 500 Hz), then the output
goes at 24 Vdc and stays at 24 Vdc until the frequency be-comes
lower than 495 Hz (that is 500 Hz – 1% of 500 Hz).
TwisTorr Medium-TMP on-board Controller User Manual / 87-901-050-01
reference quantity: frequency
threshold: 500 Hz
hysteresis: 1%
activation type: "high level"
Set point output
Technical Information
Interconnections
151/181
17
loading

Este manual también es adecuado para:

X3512-64006X3512-64016