Setup the time
1. From the "Sys Setup" screen, press the Navigation buttons to move to "Setup Time"
2. Press the power button to enter the time setup screen.
3. Short press the power button to select the item you want to change. H is Hour, M is Min-
ute, S is Second.
4. Press the navigation buttons to change the number.
5. Once the time is set, press and hold the power button to return to the previous screen.
Sound settings
This setting controls the beep during heartbeat measurements.
Use the navigation button to highlight the sound icon
(a speaker icon), and press the power button to turn the
heartbeat beep on or off.
HR Replay speed settings
The devices are able to replay heart rhythm recordings. The
current replay speed is indicated by the icon (> slow, >>
normal, >>> fast). To change the speed, use the navigation
button to highlight the speed icon, and press the power button
to change it to the desired speed.
Filter settings
The Filter settings help to reduce the effect of environmental
conditions (such as florescent lighting) that may affect
measurements. To change Filter settings, enter the "Sys Setup"
menu, and press the navigation button to highlight the Filter
icon (a wave). Press the power button to change the Filter
settings to ON or OFF.
Power-Off Timer Settings
The device automatically powers itself off after several seconds.
To change this setting, enter the "Sys Setup" screen, and press
the navigation button to highlight the Power-Off Timer (120,
60, 30, 10 seconds).
35
BewellConnect - BW-HR1 - User Manual - 042016
Delete Data
.
This screen allows you to delete recorded measurements.
Use the LEFT and RIGHT buttons to select a recording, and
then press the POWER button to delete it. If the device has
no records, the screen displays "Warning! Record is empty".
About
This screen provides device information such as software
version, serial number and other information about the
device.
Note: About the setting, use the navigation button to change the value, and the power button
to save your setting.

3. STORAGE AND MAINTENANCE

- Clean the device electrodes using a medial disinfectant swab only. Do not use water, or
substances such as benzene, gasoline, paint thinner, concentrated alcohol, or detergents.
Do not clean other parts of the device with any liquid.
- Do not drop or hit this device.
- Do not disassemble the device.
- Do not sterilize this device in an autoclave, ultraviolet sterilizer or gas sterilizer (EOG, for-
maldehyde, high density ozone, etc.).
- This device does not require calibration during the expected life cycle.
- The environment temperature for transport or storage of the packaged device is (-25°C
~70°C), and the humidity is ≤93%, no condensation.
- Avoid extreme changes in temperature and humidity. Do not use this device in locations
subject to high or low temperatures or humidity. Use at a temperature within 5°C (41°F) to
40 °C (104°F) and 15%~93% RH.
- Do not store the device in ambient conditions: exposed to direct sunlight, with high tem-
peratures and/or high humidity, that are wet or damp or where water may get on the
device, that are dusty near fires or open flames or exposed to strong vibration, or strong
electromagnetic fields.
- Dispose of the device in accordance with applicable local regulations.
- This device is designed to be compliant with the rules and regulations in locations which it
is sold and will be labeled as required.