CAIRE Spirit 300 Manual Del Usuario página 11

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 53
Setting
Off
1 (Pulse)
1.5 (Pulse)
2 (Pulse)
3 (Pulse)
4 (Pulse)
5 (Pulse)
CF (Continous)
* Trigger Level: -0.21 ± 0.11 cm H2O; Tolerance is +/- 15%; Flow
Meter Accuracy: 2% of reading or .005 SLPM; Resolution .01 LPM;
Volumetric Accuracy: 2% of reading
** The volume delivered in the 1st .40 sec @ 30 BPM is 42mL.
*** The volume delivered in the 1st .48 sec @ 25 BPM is 58.6 mL
and that delivered in the 1st .40 sec @ 30 BPM is 44mL.
WARNING: THE SETTINGS OF OTHER
MODELS OR BRANDS OF OXYGEN THERAPY
EQUIPMENT DO NOT CORRESPOND WITH
THE SETTINGS OF THE SPIRIT.
9. To verify the level of liquid oxygen in unit with
the contents indicator:
• Support the unit directly above the black
contents indicator spring scale chamber.
• Allow the unit to stabilize (i.e. not bounce up
and down), then read the exposed colored
scale for approximate liquid oxygen contents.
NOTE: If the unit is empty, only the red colored band
of the scale will be exposed. If this is the case, the unit
must be filled prior to use.
Volume
-0-
12 to 18 mL
18 to 26 mL
26 to 34 mL
40 to 50 mL
54 to 66 mL**
67 to 83 mL***
1.70 to 2.30 LPM
Condensate Pad
1. Under certain environmental conditions,
especially with continuous flow and little
movement, the Spirit may develop excess frost
around the vaporizer coil and on the case. You
may reduce this frost by tapping the unit and/or
wiping any accumulated frost off the case.
2. After the Spirit is empty and has warmed up to
room temperature, remove any moisture from
the condensation pad.
• Unzip the zipper at the base of the unit to access
the condensation pad.
• Remove the pad and wring out any absorbed
moisture. Allow the pad to dry completely
before replacing it back in the unit.
• Place dry pad back in the unit. Make sure to
completely rezip the zipper.
PN MN235-C4 C | User Manual
Spirit
11 - ENG
Tabla de contenido
loading

Este manual también es adecuado para:

Spirit 600Spirit 1200

Tabla de contenido