8
Detailed information about your system
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch Verio IQ®
System compares well with a laboratory method.
# of
# of
Slope
Intercept
Subjects
Tests
(mg/dL)
100
300
0.95
4.81
95% CI
95% CI
Std. Error
R
2
Slope
Intercept
(S
)
y.x
(mg/dL)
(mg/dL)
0.94 to
2.76 to
9.73
0.99
0.96
6.87
108