Specifications
7.1 AD101B amplifier board specifications
Type
Accuracy class (with int. calibration)
Number of trade values, acc. to
EN 45501 (R76)
Input sensitivity
Measuring range
Input signal range
Measuring signal resolution, max.
Measuring rate (depending on output
format and baud rate)
Cut-off frequency of the digital filter
(−3 dB), adjustable
Bridge excitation voltage U
(Excitation from supply voltage)
Measuring signal input, Strain gage
transducers (Full bridge)
Transducer connection
Input resistance (differential)
Transducer cable length
Interface cable length RS-232
Calibration signal
Temperature stability of the calibr. signal
Linearity error (related to full scale)
Temperature effect on
zero point (related to full scale)
measuring sensitivity (rel. to actual value)
Interface
Baud rate, adjustable
Supply voltage
Current consumption (without load cell)
Temperature range:
Nominal temperature
Service temperature
Storage temperature
Dimensions
Weight, approx.
Degree of protection according to
EN 60529 (IEC 529)
1)
Depending on the external supply voltage
Not intended for use with AED9401A, AED9501A and CANopen/DeviceNet com-
munication.
HBM
AED System Digital Transducer Electronics
d
μV/e
mV/V
bit
Hz
B
V
DC
Ω
MΩ
m
m
mV/V
ppm/°C
%
%/10K
bit/s
V
DC
mA
°C [°F]
mm
g
AD101B
0.015
6000
1
±2.0
±3.0
20 (at 1 Hz)
600 ... 4.7
40 ... 0.25
5 ... 10 (= supply voltage!)
≥40...4000
6-wire circuit
>15
≤100, in case of calibration incl. cable
≤15 (25 pol. female connector)
2±0.01 %
≤2.5
±0.01
typ. ± 0.005; max. 0.01
typ. ± 0.005; max. 0.01
RS-232
1200 ... 38400
5 ... 10
residual ripple ≤10 mV (p.p.)
≤ 80
−10...+40 [+14...+104]
−20...+60 [−4...+140]
−25...+85 [−13...185]
93 x 53 x 17
40
IP00
A1780−4.3 en/de/fr/es/it