6. Functioning
6.1.
Powering on
The device is equipped with a switch placed on the rear of the device [2]. It is activated by changing this switch
to position « I ». The light source is set to the last power used or to the value saved by the practitioner in imagyn.
6.2.
Connecting the endoscope to the camera head
Bring the dewclaws on the ring of the lens closer together until the endoscope can be inserted. Once the endos-
cope is inserted, release the two dewclaws.
The endoscope can be unlocked by simply bringing the dewclaws closer together.
6.3.

White balance

White balance is triggered by pressing and holding the [12] button on the camera head. The following procedure
should be followed:
- Once the camera is paired with the endoscope and the light source is activated, film a white surface.
- Start the white balance by pressing the corresponding button.
- Continue filming the white surface as long as the message «Processing AWB...» is present on the screen.
- Keep the button pressed until the message «AWB OK» appears on the screen. This delay avoids unin-
tentional press.
6.4.
Focus
Use the lens ring associated to the camera head to focus.
Once the endoscope is connected and the light source activated, slowly turn the ring to find a position where the
objects observed are sharp.
Focusing far enough away allows a sufficient depth of field to be obtained for the operation, thus avoiding regular
focusing.
6.5.
Other camera head buttons features
The [13] button on the camera head switches from a A value to a B value of the light source. These basic values
are set at 75% and 100% light output but can be adjusted via imagyn.
The [14] button allows:
- by short press, to launch an image capture,
- by long press, to start a video recording. Another long press stops the video recording.
6.6.
Turning off the product
To turn off the device, put the switch on the rear panel of the control unit [2] to the « O » position.
D900 700 053 B
Instructions for use: iCare imaging station
Rev 09/2018