Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/Flicker
IEC 61000-3-3
11.2. Guidance and manufacturer's declaration: electromagnetic immunity
The « CMOS camera + LED source » reference equipment has been designed for use in the electromagnetic envi-
ronment specified below. The user must ensure that it is actually used in this environment.
Immunity test
Electrostatic
± 6 kV via contact
discharges
± 8 kV via air
IEC 61000-4-2
Rapid transient
± 2 kV power lines
peaks
± 1 kV input/output lines
IEC 61000-4-4
Electric shocks
Differential mode ± 1 kV
IEC 61000-4-5
Common mode ± 2 kV
Power outages,
<5% UT - for 10 ms
short power ou-
40% UT - for 100 ms
tages and voltage
70% UT - for 500 ms
fluctuations
<5% UT - for 5 ms
IEC 61000-4-11
Magnetic field at
mains frequency
3 A/m
(50/60 Hz)
IEC 61000-4-8
Note: UT is the nominal value of the electrical voltage applied during the test.
Guidance and manufacturer's declaration: electromagnetic emissions
The « CMOS camera + LED source » reference equipment is intended for use in the electromagnetic environment
specified below. The user must ensure that it is actually used in this environment.
D900 700 053 B
Instructions for use: iCare imaging station
Compliance
This « CMOS camera + LED source » product only uses radio
Group 1
power for its subsystems. It therefore emits very low RF ener-
gy and is not likely to interfere with nearby electronic devices.
Class A
This « CMOS camera + LED source » product must be used
in all installations, other than residential installations and
Compliant
premises directly connected to the public low voltage power
distribution network intended to supply residential buildings.
Compliant
IEC 60601
Severity Level
± 6 kV
± 8 kV
± 2 kV
± 1 kV
± 1 kV
± 2 kV
5% UT 10 ms
<40% UT 100 ms
<70% UT 500 ms
<5% UT 5 s
3 A/m
Rev 09/2018
Electromagnetic environment: Guidance
Compliance
Electromagnetic environment: Guidance
Level
The floor must be made of wood, concrete
or tiles. If the floor is covered with a synthe-
tic material, the relative humidity must be at
least 30%.
The quality of the main power supply must
be the one of a typical commercial or hospi-
tal environment.
The quality of the main power supply must
be the one of a typical commercial or hospi-
tal environment. If the user of this product
must be able to continue working during
power outages, it is recommended that this
product be powered by UPS or battery.
The magnetic field at the mains frequency
must be at a characteristic level of a loca-
tion (50/60 Hz) in a typical commercial or
hospital environment.
delmont
imaging
13