Section 18 Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The CooperSurgical LEEP PRECISION Generators are intended for use in the electromagnetic environment speci-
fied below . The customer or the end user of the CooperSurgical LEEP PRECISION Generator should assure that it is
used in such an environment .
Immunity Test
Electromagnetic
discharge (ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a .c . mains voltage prior to application of the test level; in this case, 230 V .
T
IEC 60601
Test Level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/output
±1 kV for input/
lines
output lines
±1 kV differential
±1 kV differential
mode
mode
±2 kV common mode
±2 kV common mode
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 0 .5 cycle
for 0 .5 cycle
40% U
40% U
T
(60% dip in U
)
(60% dip in U
T
for 5 cycles
for 5 cycles
70% U
70% U
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
<5% U
<5% U
T
(>95% dip in U
)
(>95% dip in U
T
for 5 sec
for 5 sec
3 A/m
3 A/m
Compliance Level
Floors should be wood, concrete,
or ceramic tile . If floors are
covered with synthetic material,
the relative humidity should be at
least 30 percent .
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality should be
T
)
that of a typical commercial or
T
hospital environment .
If the user of the CooperSurgical
T
)
LEEP PRECISION Integrated
T
System requires continued
operation during power mains
interruptions, it is recommended
T
)
that the CooperSurgical
T
LEEP PRECISION Integrated
System be powered from an
uninterruptible power supply or
T
)
a battery .
T
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment .
27
Electromagnetic
Environmental Guidance