Replacement Parts - Cuda surgical Elite 300 Manual De Servicio

Ocultar thumbs Ver también para Elite 300:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

12.
Install copper clips.
13.
Remove the end cap from the front of the lamp. Use care not to touch the
window of the lamp.
14.
Apply heatsink compound to the front outside diameter of the lamp.
15.
Press the cooling ring onto the front of the lamp.
16.
Press the front heatsink assembly onto the lamp/cooling ring assembly.
Align the threaded holes in the bottom of the 2 heatsinks with each other.
17.
Secure the front heatsink onto the lamp with the copper clips.
18.
Align the threaded holes in the bottom of the heatsink with the bottom of
the lamp housing, and press the entire lamp assembly into the lamp
housing, so that when the lamp is pointing toward you, the window of the
lamp is pointing to the front panel. Install the 2 brass plugs into the bottom
of the lamp assembly and tighten.
DESCRIPTION
Replacement Lamp Cartridge
EMI Filter
300-Watt Xenon Power Supply
Hour meter
Circuit Breaker
IR Filter Assembly
Power Input Module
Cooling Fan
Reset Switch Assembly
Xenon Lamp Rebuild Kit
Shutter Assembly
Lamp Base Assembly
For parts not listed here, call our Technical Service Department listed at the front of the
manual.
LIT194 CUDA SURGICAL
(English)

REPLACEMENT PARTS

PART NUMBER
SSR-001
SSR-002
SSR-003
SSR-004
SSR-005
SSR-006
SSR-007
SSR-010
SSR-011
SSR-012
SSR-013
SSR-014
Page 10 of 64
Rev.
Tabla de contenido
loading

Tabla de contenido