Circuit Breaker Replacement; Shutter Replacement - Cuda surgical Elite 300 Manual De Servicio

Ocultar thumbs Ver también para Elite 300:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

1.
Disconnect the illuminator from the power source.
2.
Remove the 8 cover screws on the sides of the unit.
3.
Open lamp access door and lift the top cover straight up and off the unit.
Set aside.
4.
Using an ohmmeter, verify which circuit breaker is defective.
5.
Remove the defective circuit breaker and replace with the new one.
6.
Replace the top cover making sure the lamp access door is over the lamp
cartridge.
7.
Replace the 8 cover screws and secure.
1.
Disconnect the illuminator from the power source.
2. Remove the ten cover screws on the sides and the two on top of the unit.
3. Remove O-ring that engages knob with shutter.
4. Remove the E-external retaining ring holding the shutter assembly to the
front panel.
5. Slide off the shutter assembly from mounting stud.
6. Replace the shutter assembly with the new one. Protruding set screw in
shutter assembly must go up between two pressed in pins on sheet
metal cabinet casing. Re-install E-external retaining ring.
7. Re-install O-ring. Turn intensity knob left and right until shutter is in synced
with intensity knob marking.
8. Replace the top cover making sure the lamp access door is over the lamp
cartridge.
9. Replace the twelve cover screws and secure.
LIT194 CUDA SURGICAL
(English)

CIRCUIT BREAKER REPLACEMENT

SHUTTER REPLACEMENT

Rev.
Page 6 of 64