Alarm Indicators and Screen Symbols

Visual, Audio
Indicators
A sequence of
3 short
beeps
+
1 long beep
that repeats every
4 seconds
A sequence of
3 long beeps
+
1 short beep
that repeats every
4 seconds
Description
No Breath Alarm
This alarm occurs
when a breath is not
detected for a period
of 2 minutes or more�
It can be silenced by
pressing the Alarm
symbol� This alarm
resets itself as soon as
a breath is detected� If
no breath is detected
after approximately
30 minutes, the device
shuts down to conserve
power� (This alarm is
inactive for ten minutes
after startup�)
Low Oxygen
Concentration Alarm
This alarm occurs when
the device is delivering
a lower concentration of
oxygen than specified�
This alarm can be
silenced by pressing the
Alarm symbol�
Note: This alarm occurs
when the internal O
sensor reads < 82% O
This alarm will turn off
if the system is able to
recover from the fault
and achieve >84% O
SimplyGo User Manual
What To Do
Check the connection
from the cannula to
the device� Make sure
the nasal cannula is
properly positioned
on your face and that
you are breathing
through your nose�
Make sure the
cannula tubing is not
kinked or obstructed�
Change to another
source of oxygen
and contact your
equipment provider�