Setting Up Dotting Recorder - Endress+Hauser Chroma-Log Serie Manual De Utilización

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

3.7 Setting up dotting recorder

3.7.1 Pull chassis from casing (pictures see 3.6.1)
Completely open the recorder front door, remove the
cassette. Lift the blocking spring
part of the cassette area); Grip the chassis by the holes
in the separation panel
(in the upper part of the
Ë
cassette area) and pull the chassis out up to the stop.
3.7.2 Operating components
All operating components are fitted to the top printed
circuit board and are covered by a steel cover. Printed
on this cover plate is a short form operating instruction.
Operating elements
3.7.3 Setting up measurement ranges
Selection of measurement range is done using DIP
switches 3 and 4 on the PCB. Each channel can be set
up individually.
3.7.4 Settings DIP switches 3 and 4
In the example all channels are set to the same range.
Measurement range DIP
switches
3 Installation/Setting up
(in the lower right
Ê
39