6. OPERATING GUIDE

6.1 Application method

A
a) Put the suitable probe into the jack on the right side of the oximeter. (The probe is limited to be pro-
duced by our company; never replace it with the similar ones by other manufacturers).
b) Put the finger into the probe.
c) Press the "power on/off button" long , until the device turns on.
d) Do not shake the finger and keep the patient in a stable state during the process.
e) The data can be read directly from the screen on the measuring interface.
Fingernails and the luminescent tube should be on the same side.
If the alarm function is on, the device will provide medium-priority alarm signal when probe or
finger is out. Intermittent alarm will occur and the user interface presents "FINGER OUT".
Medium priority indicating that prompt operator response is required.
B. PAUSE ALARM
a) Alarm including the alarm of measure data's going beyond the limits, the alarm of low-voltage, the
alarm of probe or finger's out of position.
b) When alarm is on, press the "alarm pause button" can pause the alarm, it can renew alarm in about
60s, and if pressing the "alarm pause button" Again with in 60s, it can renew alarm.
c) If you want to turn off the alarm for good, you should enter the menu for operation.
C. REVIEW INTERFACE
a) On the measuring interface, press "up button" to enter the Review Interface 1 directly, as shown in
figure 5:
b) In review interface, press "menu button" to switch between Review Interface 1 and Review In-
terface 2; press "Down button" to enter the review interface for next hour or last hour; In Review
Interface 1, press "left button" or "right button" can move the trend graph for storage data, When
the trend graph cannot be moved any more, the sign —— "<-" or "->" shown under the LCD screen
will disappear; in Review Interface 2, press "left button" or "right button" can move the arrow; Press
"up button" to exit the review Interface.
c) In Review Interface 1, can observe the trend waveform composed by storage data, each screen
can show storage data for 114 seconds, the yellow line shows the SpO
shows the PR trend waveform, the time underside shows the starting time of displaying the date in
the screen, the middle "+" and "-" underside the screen means the operation direction of the "Down
button". Press "right button", it will show "+" in the position, then press "Down button" to enter next
hour; Press "left button", it will show "-" in the position, then press "Down button" to enter last hour.
d) The Review Interface 2 shown based on Review Interface 1, the stored SpO
in each second can be observed here, the underside date from left to right marks time, SpO
PR value. when the stored data exceeds the upper and lower limit setted by user, the relevant value
will turn green.
Figure 5. Review Interface 1
ENGLISH
trend waveform, the red line
2
value and PR value
2
value,
2