CLEanInG anD DISInFECTInG InSTRUCTIOnS
CLEANING ZEON
Zeon® Endeavour™ is a LED headlight manufactured with high-quality components; a precision
optical instrument that must be cleaned and handled with the greatest care . Remove contaminants
using a soft, clean, dry cloth in order to avoid scratching the headlight .
!
CAUTION
Do not pull on the Zeon
Clean all lens surfaces with an optical lens cleaning cloth that is intended for use with coated lens
surfaces . you may purchase these cleaning cloths from Orascoptic or at major retail chains, optical
stores, and drug stores .
If your loupes are heavily soiled, first remove Zeon® Endeavour™ headlight and cable, and then
clean your loupes. Be sure to dry your loupes thoroughly before reinstalling Endeavour.
!
CAUTION
Do not submerge Zeon® Endeavour™ components, including the LED headlight
and/or the battery pack, in any liquid or an ultrasonic cleaner, because damage may
result . Endeavour components are not waterproof .
Do not use gasoline, cleaning solutions, or aggressive solvents for cleaning .
DISINFECTING ZEON
Disinfect Zeon
Endeavour™ components, including the battery charger, using CaviWipes
®
disinfection cloth . Use disinfection cloths only for severe contamination .
!
CAUTION
Do not use an autoclave, chemiclave, glutaraldehyde, iodophor or any
other sterilization method, or damage may result.
Disinfection spray must not be used under any circumstances
to avoid severely damaging the Zeon® Endeavour™ components, including
the battery charger .
Only use Caviwipes
- is suitable for alcohol-sensitive surfaces and
for medical facilities, such as a dentist's chair .
- contains no type Aldehyde or phenol
- has a broad-spectrum effect against bacteria,
fungi, and viruses .
If there are any questions, call Customer Service at the number listed on the back cover of this
manual .
OPERATING INSTRUCTIONS
Zeon
endeavour™ Led Light SyStem
®
ENDEAvOUR™ COMPONENTS
®
Endeavour™ cables during cleaning!
®
ENDEAvOUR™ COMPONENTS
®
or a disinfection cloth that:
®
9