Technical Data - Weinmann OXYMAT 3 Descripción Del Aparato E Instrucciones

Ocultar thumbs Ver también para OXYMAT 3:
128

12. Technical Data

Product category according to
93/42/EEC
Dimensions WxHxD in mm
Weight
Temperature range
– Operation
– Storage
Power supply
Power output / current consumption 360 W / 1.6 A
Mains fuse DIN EN 60127-2
Classification acc. to EN 60601-1
– Protection against electric shock
– Degree of protection against
electric shock
Electromagnetic compatibility
– Radio interference suppression
– Radio interference resistance
Sound pressure level
O
output (values after 10 min.
2
operation at + 20 °C,
50% relative humidity and
1013 mbar)
O
output at 2000 m above msl
2
Oxygen outlet pressure
Flow rate (read at top of ball)
OSCI O
status indicator as per
2
DIN EN ISO 8359.51.5
Warning at < 82 % by vol. of O
Fault warning
Actuating pressure of pressure
relief mechanism
Servicing
English
OXYMAT 3
IIa
400 x 700 x 350
approx. 20 kg
+10 to +40 °C
- 20 to +70 °C
230V, 50 Hz
T 2.5 A H 250V
Class 2
Type B
EN 55011 (VDE
0875 Pt11)
IEC 1000-4 Parts
2-6 and 11
≤ 40 dB(A)
(from serial num-
ber 20.000)
95 +1/-3 vol% O
2
at 1–4 l/min
90 ±3 vol% O
at
2
5 l/min
90 ±3 vol% O
at
2
5 l/min
60 kpa
0.5 – 5.5 l/min
0.2 – 3.0 l/min
present
2
Visual and acoustic
300 kPa
at least once a
year or every
5000 operating
hours

Capítulos