Chapter 4,
Calibrating the Altair Single Gas Detector
•
The Altair Single Gas detector is designed to be a maintenance-
free instrument, but it must be calibrated if it does not pass a Bump
Test or if calibration is required by local procedures.
•
For oxygen instruments, perform a calibration if:
•
there are changes in barometric pressure
(altitude changes)
•
there are extreme changes in ambient temperature and
humidity (see TABLE 6-2, "Instrument Specifications").
•
the instrument does not pass a bump test.
•
For toxic instruments (CO and H 2 S), perform a calibration if any of
the following occur:
•
physical shock
•
extended use in extreme temperatures
•
high concentration exposure
•
the instrument does not pass a bump test.
4-1