5. TECHNICAL DATA
The technical data is valid under the following conditions:
Indoor use only
Powered by external approved LPS AC/DC adaptor
Altitude: Up to 2000 m
Operating temperature range: 5 ° C to 40 ° C
Specified Temperature range: 10° C to 30° C
Humidity: maximum relative humidity 80% for temperatures up to 30° C, decreasing linearly to
50% relative humidity at 40° C
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation category II
Pollution degree: 2
PX85
Model
Capacity (g)
Readability d (g)
0.00001
Repeatability
0.00002
(std. dev.) (g)
Linearity (g)
± 0.0001
Power Supply
Model
Capacity (g)
Readability d
(g)
Repeatability
(std.
dev.)
(g)
Linearity (g)
Power
Supply
PX225D
82/220
82
0.00001/0.0001
0.00002/0.0001
±0.0001
Power input: 100-240V ~ 200mA 50-60Hz 12-18VA
PX523
PX1602
PX523/E
PX1602/E PX2202/E PX3202/E PX4202/E PX2201/E PX4201/E
520
1600
0.001
0.01
0.001
0.01
±0.002
±0.02
Power input: 100-240V ~ 200mA 50-60Hz 12-18VA
PX84
PX124
PX84/E
PX124/E
82
120
0.0001
0.0001
0.0001
0.0001
±0.0002
±0.0002
Power output: 12 VDC 0.5A
PX2202
PX3202
2200
3200
0.01
0.01
0.01
0.01
±0.02
±0.02
Power output: 12 VDC 0.5A
PX224
PX163
PX224/E
PX163/E
220
160
0.0001
0.001
0.0001
0.001
±0.0002
±0.002
PX4202
PX2201
4200
2200
0.01
0.1
0.01
0.1
±0.02
±0.2
EN-5
PX323
PX323/E
320
0.001
0.001
±0.002
PX4201
4200
0.1
0.1
±0.2