Defender 3000 Series Floor Scale
MAINTENANCE
Daily Checks
Check that the rocker feet are properly seated in the holes of each load cell.
Check that the 4 feet are in contact with the floor.
Check that the scale is free from debris, especially under the edges of the scale.
Check that there is no damage to the connecting cable.
Cleaning
Attention: Do not use solvents, harsh chemicals, ammonia or abrasive cleaning agents.
The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
Maintenance of the floor scale is limited to regular cleaning.

TECHNICAL DATA

Equipment Ratings:
Altitude:
Operating temperature:
Humidity:
Pollution degree:
Specifications
Model
Capacity (kg)
Readability (kg)
Dimensions
Material
Feet
Load Cells
Load Cell Safe Overload
Limit
Load Cell Output Impedance
Load Cell Excitation Voltage
Junction Box
Net Weight
Model
Capacity (kg)
Readability (kg)
Dimensions
Material
Feet
Load Cells
Load Cell Safe Overload
Limit
Load Cell Output Impedance
Load Cell Excitation Voltage
Junction Box
Net Weight
Cleaning i-DFxxxB1x models, (Painted Steel version) in a dry environment:
- Wipe surfaces with a damp cloth.
- Use only common household cleaning agents.
- Never use water spray or high pressure cleaners.
Cleaning the i-DFxxxC1x models, (Stainless Steel version) in a wet environment:
- Water jets up to 60°C
- Use only common household cleaning agents.
- Ensure that corrosive substances are removed at regular intervals.
All models, when using optional Ramps or Foot Plates:
- Keep the air gap at the edge of the Floor Scale free of debris.
2,000m
-10°C to 40°C
Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to 50%
relative humidity at 40°C.
2
TABLE 3. FLOOR SCALE SPECIFICATIONS
i-DF600B1R i-DF1500B1R
600
1500
0.2
0.5
1m x 1m
1m x 1m
Painted Carbon Steel, treaded surface, fixed top plate
4 x swivel design, rubber bottom, adjustable height
Rear panel access, IP65, Stainless Steel, with 5 m Cable
112 kg
TABLE 4. FLOOR SCALE SPECIFICATIONS
i-DF1500B1RL
1500
0.5
1m x 1.25m
Painted Carbon Steel, treaded surface, fixed top plate
4 x swivel design, rubber bottom, adjustable height
Rear panel access, IP65, Stainless Steel, with 5 m Cable
120 kg
i-DF1500B1L
i-DF3000B1L i-DF1500B1X i-DF3000B1X
1500
3000
0.5
1.25m x 1.25m 1.25m x 1.25m 1.5m x 1.5m
4 x Shear Beam, Alloy Steel, OIML C3
120% of Emax
5V ~15V, DC/AC
138 kg
i-DF3000B1RL
i-DF1500B1LX
3000
1
1m x 1.25m
1.25m x 1.5m
4 x Shear Beam, Alloy Steel, OIML C3
120% of Emax
5V ~15V, DC/AC
1500
1
0.5
174 kg
i-DF3000B1LX
1500
3000
0.5
1.25m x 1.5m
154 kg
EN-3
3000
1
1.5m x 1.5m
1