Appendix E Error Codes; Table E-1. Error Codes - Ohio Medical Corporation MiniOX 3000 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

MiniOX
3000 Oxygen Monitor
®
Error Codes
The MiniOX
3000 Oxygen Monitor performs diagnostic
®
tests to detect errors that could cause unreliable
monitor operation. These tests are performed:
when a battery is installed (power-up)
when the monitor is turned ON
during routine monitoring
During self-test or any time the monitor detects an
operational error, the MiniOX
ceases operation
display "ERR" and an error code
activates visual and audible alarms
Error alarms continue until:
ERROR DISPLAY
01
02
03
04
05
06
07
08
*Press SET. If error reoccurs, call Ohio Medical Corporation Repair
**Replace battery. If error reoccurs, call Ohio Medical Corporation Repair
3000 Oxygen Monitor:
®

Table E-1. Error Codes

System error
Random Access Memory (RAM) error
Read Only Memory (ROM) error
Analog error
Timing error
Keypad error*
Battery error **
LCD error
711449 (Rev.3) 02/2016
1. The monitor is shut off (press l/0)
2. The monitor is reset and error does not reoccur
(press set)
3. The battery is disconnected
If an error reoccurs, the monitor is inoperative and must
not be used; call Ohio Medical Corporation Repair at:
1-847-855-0800
The error code that appears on the display corresponds
to a specific failure. The audible alarm is a beep that
corresponds to this code (e.g., "ERR 3" triggers a
three-beep alarm).
CORRESPONDING ERROR
Appendix E, Error Codes
E-1
Tabla de contenido
loading

Tabla de contenido