MiniOX
3000 Oxygen Monitor
®
Electromagnetic Compatibility (EMC)
Ensure that the specified electromagnetic environment and separation distances for the MiniOX
Oxygen Monitor are observed according to the tables below. Misuse may result in failure to perform as
designed, including incorrect/erratic readings.
Guidance and manufacturer's declaration - electromagnetic emissions
The MiniOX
3000 is intended for use in the electromagnetic environment specified below.The customer or the
®
user of the MiniOX
3000 should ensure that it is used in such an environment.
®
Emissions test
Compliance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Not applicable
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The MiniOX
3000 is intended for use in the electromagnetic environment specified below. The customer or the
®
user of the MiniOX
3000 should ensure that it is used in such an environment.
®
Immunity test
Electrostatic
+6 kV contact
discharge (ESD)
+8 kV air
IEC 61000-4-2
Electrical fast
+2 kV for power supply
transient/burst
lines
IEC 61000-4-4
+1 kV for input/output lines
Surge
+1 kV line(s) to line(s)
IEC 61000-4-5
+2 kV line(s) to earth
Voltage dips, short
Not applicable
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note:U
is the a.c. mains voltage prior to application of the test level.
T
WARNING
Electromagnetic environment – guidance
The MiniOX
®
Group 1
fore, its RF emissions are very low and are not likely to cause any inter-
ference in nearby electronic equipment.
The MiniOX
®
domestic establishments and those directly connected to the public low-
Class B
voltage power supply network that supplies buildings used for domestic
purposes.
Not applicable
Not applicable
IEC 60601
test level
+6 kV contact
+8 kV air
Not applicable
Not applicable
Not applicable
3 A/m
711449 (Rev.3) 02/2016
3000 uses RF energy only for its internal function.There-
3000 is suitable for use in all establishments including
Compliance
Electromagnetic environment –
level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Not applicable
Not applicable
Not applicable
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.
Appendix A, Specifications
3000
®
guidance
A-2