MiniOX
3000 Oxygen Monitor
®
Section 4
Maintenance and Care
CAUTION
Use only genuine Ohio Medical Corporation
replacement parts when performing any
maintenance procedures included in this
manual. Failure to do so may seriously
impair the monitor's performance.
Repair or alteration of the MiniOX
Oxygen Monitor beyond the scope of the
maintenance instructions or by anyone other
than an authorized Ohio Medical Corporation
service person could cause the product to
fail to perform as designed.
This section describes the following general
maintenance and care procedures for the MiniOX
3000 Oxygen Monitor:
•
Battery Replacement
•
Sensor Replacement
•
Deflector Replacement
•
Cable Replacement
•
Cleaning
Battery & Battery Cover Replacement
The MiniOX
standard 9-volt alkaline battery. To replace the battery:
1. Verify that the monitor is turned OFF. The display
should be blank.
2. Pull out the support stand from the back of the
case.
3000
®
3. Unscrew the two screws on the battery cover in
the back of the instrument and remove cover.
See Figure 4-1.
4. Remove the battery from the case and unsnap the
battery from the battery holder.
NOTE: To ensure proper start-up, wait at least
®
5. Snap the terminal of the new battery into the
battery holder.
6. Install the battery cover and screw into place.
Make sure that the battery cover is properly
seated and flat on the back of the MiniOX
Oxygen Monitor case.
7. Recalibrate the monitor. Reset the low and high
alarms, if desired.
Figure 4-1
Battery Cover Removal
711449 (Rev.3) 02/2016
Section 4, Maintenance and Care
3000 Oxygen Monitor requires one
®
45 seconds before connecting the
fresh battery to the battery connector.
3000
®
4-1