MiniOX
3000 Oxygen Monitor
®
portable and mobile RF communications equipment and the MiniOX
The MiniOX
3000 is intended for use in an electromagnetic environment in which radiated RF disturbances
®
are controlled. The customer or the user of the MiniOX
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the MiniOX
3000 as recommended below, according to the maximum output power of the communications
®
equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
Appendix B
Recommended separation distances between
®
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 2.9 √P
0.29
0.92
2.9
9.2
29
711449 (Rev.3) 02/2016
3000 can help prevent electromagnetic interference
(m)
80 MHz to 800 MHz
d = 0.18 √P
0.018
0.057
0.18
0.57
1.8
Appendix A, Specifications
3000
®
800 MHz to 2.5 GHz
d = 0.35 √P
0.035
0.11
0.35
1.1
3.5
A-4