Observe the safety notices in the section "Information on proximity to certain areas" (see
page 23).
Electromagnetic emissions
Interference meas
urements
HF emissions accord
ing to CISPR 11
Harmonics according
to IEC 61000-3-2
Voltage fluctu
ations/flicker accord
ing to IEC 61000-3-3
Electromagnetic interference immunity
Phenomenon
Electrostatic dis
charge
High-frequency elec
tromagnetic fields
Magnetic fields with
rated power frequen
cies
Electrical fast transi
ents/bursts
Surges
Line against line
Conducted interfer
ence induced by high-
frequency fields
Voltage drops
Voltage interruptions
Compliance
Group 1/class B
Not applicable –
power below 75 W
Product meets the
requirements of the
standard.
EMC basic standard
or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Electromagnetic environment directive
The product uses HF energy exclusively for its
internal functioning. Its HF emissions are
therefore very low, and interference with neigh
bouring electronic devices is unlikely.
Interference immunity test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
30 A/m
50 Hz or 60 Hz
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
3 V
0.15 MHz to 80 MHz
6 V in ISM and amateur frequency bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
0% U
; 1/2 period
T
At 0, 45, 90, 135, 180, 225, 270 and 315
degrees
0% U
; 1 period
T
and
70% U
; 25/30 periods
T
Single phase: at 0 degrees
0% U
; 250/300 periods
T
–
–
35