Supplement to the instructions for use
Movita/Movita lift with part number 9052693,
edition: 2 – 2012-08 and higher
Forta lift with part number 9052715,
edition: 1 – 2010-09 and higher
WARNING
Risk of operating errors and incorrect use
Strictly follow the instructions for use of the
connected devices and this supplement to the
instructions for use.
Keep this supplement with the instructions for
use of the medical device.
This supplement updates the information in the
instructions for use of the medical device in the
following chapters:
Overview
Symbols
Symbol
Description
Warning! Strictly follow these
instructions for use
Attention! (safety sign)
Ceiling-mounted version of
Perseus
Supplement to the instructions for use
Docking or undocking of Perseus
Docking, undocking, and operation
of other Dräger devices, e.g.,
Primus or Mova Cart
Operation
WARNING
Risk of personal injury or damage to the
device
Safety functions of the ceiling supply unit can
be temporarily disabled with the keyswitch
key.
After the device has been docked or undocked
by service personnel, make sure that the key
is not left in the keyswitch but instead is kept
in a secure place.
5