Verificación Del Ajuste Del Cero Con Prolink Iii - Emerson Micro Motion 1700 Manual De Configuración Y Uso

Transmisores con salidas analógicas
Ocultar thumbs Ver también para Micro Motion 1700:
Tabla de contenido
Requisitos posteriores
Restore normal flow through the sensor by opening the valves.
2.6.2
Verificación del ajuste del cero con ProLink III
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
Importante
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless
one of the following is true:
Prerrequisitos
ProLink III v1.0 con Parche de compilación 31 o versión posterior
Importante
Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem,
then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a low-
severity alarm is active.
Procedimiento
1.
2.
3.
Manual de configuración y uso
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
d. If it fails again, zero the flowmeter.
For instructions on zeroing the flowmeter, see
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Prepare the flowmeter:
a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
Seleccione Herramientas del dispositivo > Calibración del dispositivo > Verificación y calibración
de ajuste del cero > Verificar ajuste del cero y espere hasta que se complete el
procedimiento.
If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
Ajuste del cero del medidor de
Inicio rápido
caudal.
13
Tabla de contenido
loading

Tabla de contenido