receptacle on the bottom of the unit (see figure 2, letter B). The three button handpiece is designed to give the
doctor complete fingertip control of the power output settings.
5. Slide the standard electrode into the handpiece until it is firmly seated (see figure 2, letter C).
The handpiece will accept most standard
6. Slide the handpiece from above into the holder on the right side of the unit.
7. Turn the unit power on utilizing the switch on the right side panel of the unit (see figure 2, letter D).
8. Set the power output either by using the dial on the front of the unit (see figure 2, letter E) or on Aaron 940™
unit only by the up and down buttons on the handpiece (see figure 2, letter J). When power level adjustment is
being made by the handpiece an audible tone will sound to indicate that the power level has been changed.
Depressing and holding the up or down buttons will cause the power settings to change more rapidly for quick
adjustment of the output settings. Power output is displayed in ".1" watt increments below 10 watts and in
whole numbers from 10 to 40 watts.
NOTICE :
The output settings can not be adjusted when the unit is being activated.
9. To activate the unit, remove the handpiece from the holder. Place the handpiece in the desired position and
depress the activation button (see figure 2, letter K). When the unit is activated, an audible tone is sounded and
the blue active light will illuminate (see figure 2, letter F).
10. To use the optional grounding plate with cord (A802EU), insert the plug of the cord into the grounding
plate output (see figure 2, letter G) and connect the other end into the grounding plate. The plate should be
placed underneath the patient at a point where the entire plate is covered by bare skin.
11. To use the optional bipolar cord (A827EU), insert the plugs into the bipolar outputs (see figure 2, letter H).
The cord is then plugged into the forceps. A sliding gate behind the monopolar and bipolar outputs prevents
the user from using both simultaneously.
12. The optional footswitch is plugged into the footswitch output and placed on the floor (see figure 2, letter I).
The footswitch can be used with monopolar procedures and must be used with bipolar procedures.
13. When the procedure is completed, turn the unit off utilizing the switch on the right side panel of the unit.
14. Return the handpiece to the holder on the right side of the unit and remove the electrode. The electrode
should be disposed of after each procedure. If contamination has occurred to the handpiece, the handpiece
should also be sterilized.
15. Adjustment of the audible tone is achieved by a switch located on the rear of the unit (see figure 3). Two
tone choices are available, high and low. A small screwdriver will be necessary to make the adjustment.
Figure 3
8
3
/