ENTER
16. Press
A reverse font (light background
with dark characters) will appear
to indicate that the reading can be
adjusted.
17. Press
the concentration of the standard,
1.00 in the example. Use the value
on the turbidity standard bottle label
as the target. Note: The allowable
adjustment is ±25%. If the allowable
adjustment limit is reached,
"overrange" wil be displayed.
ENTER
18. Press
menu choices will be off ered, Set
Calibration and Factory Setting.
ENTER
19. Press
and save the calibration. Press
or
Factory Setting to revert to the
factory calibration. The meter will
momentarily display Storing... and
return to the Turbidity-Without Blank
menu. The calibration has now been
saved and the meter can be used for
testing. The standard should never be
poured from the tube back into the
bottle.
NOTE: For the greatest accuracy during the calibration procedure, be sure that after the
meter is blanked and the blank is scanned as a sample, the reading is 0.00. If not, reblank
the meter and scan the blank again until it reads 0.00. When scanning the calibration
standards as the sample, scan the calibration standard three times removing the tube
from the chamber after each scan and reinserting the tube in the chamber with the same
orientation. The readings should be consistent. Use the last consistent reading to calibrate
the meter. If the readings are not consistent, avoid using an aberrant reading to calibrate
the meter.
to select Calibrate.
or
to scroll to
to select Calibrate. Two
to select Set Calibration
to scroll and select
Turbidity WB
0.99
Scan Sample
Calibrate
12:00:00
001/500
Turbidity WB (N)
1.00
Scan Sample
Calibrate
12:00:00
001/500
Calibrate Menu (N)
1.00
Set Calibration
Factory Setting
12:00:00
001/500
Turbidity WB
Scan Blank
Scan Sample
12:00:00
001/500
31
NTU
NTU