Perform a more thorough Hard Reset as follows:
1. Power down the Gage and wait 5 seconds.
2. Simultaneously press and hold the (+) and
until the Reset symbol
This returns the Gage to a known, "out-of-the-box" condition.
It performs the same function as a menu Reset with the addition of:
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following status:
Bluetooth Smart = OFF
Test Time = 1 second
Sound = ON
Backlight = Normal
Flip Display = Normal
NOTE:
Date,Time and WiFi settings are not affected by either Reset.
Test Time
The PosiTector SHD has an on-screen timer to measure the
hardness of a material after a given period of time as defined
by international standards, typically 1, 3 or 15 seconds.
Use the (-) and (+) buttons to adjust the test duration. When taking
a reading, the timer starts automatically once the presser foot is on
the surface and the symbol appears on the display.
Auto Ign. 20/90
According to most international standards, readings below 20 and
above 90 are not considered reliable. When enabled, measurement
results outside 20/90 will emit a low tone and will not be included in
Memory or Statistics.
NOTE:
If values above 90 Shore A are measured, the PosiTector
SHD-D probe should be used. If values below 20 Shore D are
measured, use the PosiTector SHD-A probe.
appears.
(Auto Ignore)
5
center buttons
Auto Sync .net = ON
USB Drive= ON
White on Black = ON
Language = English
Battery Type = Alkaline