Hach DR300 Manual Del Usuario página 19

Ocultar thumbs Ver también para DR300:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
19. Immediately empty and rinse the sample cell. Rinse the sample cell
and cap three times with deionized water (or distilled water).
Note: As an alternative, tap water can be used to rinse the sample cell if the
concentration of the parameter in the tap water is less than the samples
measured.
20. Push and hold
Enter a calibration curve with the keypad
At least two data pairs are necessary to enter a user-prepared
calibration curve. A concentration value and the absorbance value for
the given concentration is necessary for each data pair. A maximum of
10 data pairs can be entered.
1. Push
to set the instrument to the range to calibrate (e.g., LR or
HR).
2. Push and hold
3. Push
until "USEr" shows, then push
4. Push
until "Edit" shows, then push
5. When "S0" shows on the display, push
6. Enter the first data pair.
The first data pair is S0 (concentration value) and A0 (absorbance
value).
• Push
or
• Push
to go to the next digit.
• Push
to go to the previous digit.
7. Do steps
5
8. The calibration is completed with two data pairs. If additional data
pairs are necessary for calibration:
a. When "Add" shows, push
b. Do steps
9. Push and hold
for 3 seconds to go back to measurement mode.
for 3 seconds to enter menu mode.
to change the number that flashes.
and
6
again to enter the second data pair (S1 and A1).
5
and
6
again to enter more data pairs.
for 3 seconds to go back to measurement mode.
.
.
.
.
English 19

Capítulos

Tabla de contenido
loading

Tabla de contenido