Notes
1. The device is manufactured according to applicable standards and complies with all requirements on
protection relating to electromagnetic compatibility; special precautions should however be taken for
EMC and the device should be installed and operated in compliance with the information on EMC in
this leafl et.
2. Portable and mobile radio communication equipment may affect operation of the electromedical de-
vice. When electromagnetic disturbance occurs, repeat arterial blood pressure readings if values are
uncertain.
TABLE 201 – electromagnetic emissions
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The CS 530 is intended for use in the electromagnetic environment specifi ed below. The customer or user of
the CS 530 should ensure the device is used in this environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Complies
Electromagnetic environment - guidance
Group 1
The CS 530 monitor only uses RF energy for its internal function,
therefore its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The CS 530 monitor is suitable for use in all environments
including domestic ones and those directly connected to the public
power supply network that supplies buildings used for domestic
purposes.
27