INTERCONNECTIONS
Input power connector (Line type)
The following figure show the input power connec-
tor.
P3 - Vent
This is a dedicated 24 Vdc connector to control the
optional vent valve.
P5 – External fan
This is a dedicated 24 Vdc connector to supply an
external fan.
J5 - IN-OUT
This connector carries all the input and output sig-
nals to remote control the Mini-TASK.
It is a 15-pins D type connector; the available sig-
nals are detailed in the table, the following para-
graphs describe the signal characteristics and use.
PIN N.
1
START/STOP (+)
2
START/STOP (-)
3
INTERLOCK (+)
4
INTERLOCK (-)
5
SPEED SETTING (+)
6
SPEED SETTING (-)
7
SOFT START (+)
8
SOFT START (-)
9
+24 Vdc
10
SPARE
11
PROGRAMMABLE SET POINT
12
SPARE
13
FAULT OUTPUT
14
PROGRAMMABLE ANALOG
SIGNAL (+)
15
•
GROUND (24 Vdc)
•
PROGRAMMABLE ANALOG
SIGNAL (-)
Signal description
START/STOP: input signal to start or stop the
pump. Without the supplied cover connector, if you
want to stop the turbo-pump from outside, you
have to connect the START/STOP (+) signal to the
+24 Vdc pin and the START/STOP (-) signal to the
GROUND pin (this is the default configuration of
the cover connector).
INTERLOCK: input signal to control the pump ro-
tation. With the supplied cover connector the IN-
TERLOCK (+) signal is connected to the +24 Vdc
pin and the INTERLOCK (-)signal to the GROUND
pin: in this condition the pump is stopped.