Analogue Inputs - Ecolab Versatronic Instrucciones De Servicio

Ocultar thumbs Ver también para Versatronic:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Go to: On-line parameters è Verify / test (double-click) è Hardware / software tab
Calibration constants
Displays for each analogue input and output characteristic curve data (e.g., calibration of
analytical inputs).
Go to: On-line parameters è Verify / test (double-click) è Invariable constants tab

Analogue inputs

Here, the analogue inputs of the device can be tested for function. For a correct test of an
analogue input, the mains frequency of the device connection must be set in addition to
the selection of the analogue input and signal type. A wiring diagram for the test signal is
displayed by pressing the "Test" button. Ensure that the input has a valid test signal in
accordance with the connection diagram. The test run is started by confirming the
connection plan. You can view the readings of each analogue input in the actual value and
the reading table. To stop the test run, click "Stop".
Go to: On-line parameters è Verify / test (double-click) è Analogue inputs tab
Fig. 141: Analogue inputs
(1) The "Stop" button stops a test run
(2) Process value shows the current analogue value of the
input
(3) The "Test" button starts a test run
(4) Table of measurements lists the readings over the time
period
(5) Set "Supply"by chosing the mains frequency of the
device connection
(6) Selection of the analogue input that is to be tested
(7) Selection of the signal type that is to be tested
239
PC setup program
417102269 Rev. 2-08.2020
Tabla de contenido
loading

Este manual también es adecuado para:

255201

Tabla de contenido