Analogue Signal Alerts - Ecolab Versatronic Instrucciones De Servicio

Ocultar thumbs Ver también para Versatronic:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Device menu

10.4.1.1 Analogue signal alerts

View hold behaviour for threshold monitoring:
Device menu è Configuration è Limit monitoring è Limit monitoring X è Alert
behaviour
View hold behaviour for temperature monitoring:
Device menu è Configuration è Analogue inputs è IN X temperature input X è Alert
behaviour
View hold behaviour for universal inputs:
Device menu è Configuration è Analogue inputs è IN X universal input X è Alert
behaviour
View hold behaviour for analysis inputs:
Device menu è Configuration è Analogue inputs è IN X universal input X è Alert
behaviour
Configuration
item
Binary signal for
Hold
Hold behaviour
Behaviour during
calibration
Error Handling
Call Alarm Configuration Limit monitoring:
Device menu è Configuration è Limit monitoring è Limit monitoring X è alarm alarm X
View alarm configuration for universal inputs:
Device menu è Configuration è Analogue inputs è IN X universal input X è Alarm
behaviour è Alarm X
View alarm configuration for analysis inputs:
Device menu è Configuration è Analogue inputs è IN X measurement X è Alarm
behaviour è Alarm X
Configuration item Selection / option
Alarm type
Alarm / event list
Alarm text
Threshold
417102269 Rev. 2-08.2020
Selection /
option
Explanation
Selection from
Binary signal for activating the hold function
the binary
With the hold function enabled, the alarm assumes the
selector
state defined in the setting "Hold behaviour".
Inactive
Set the alarm state with the hold function enabled, when the
Active
input is calibrated of in an error case (measurement range
Frozen
exceeded / not reached)
inactive
Inactive: Alarm is suppressed
active
active: Alarm is enforced
frozen normal
inactive
Frozen: Alarm condition is maintained regardless of
active
changes in the alarm condition
frozen
Normal: Alarm according to alarm condition
Inactive
Min. Alarm
Max. alarm Alarm
window
Inverted alarm window
OFF
Alarm
Event
Up to 21 characters of
text
-99999 to +99999
146
Explanation
4 alarm types (comparable functions) can be
selected to monitor readings for threshold
violations.
See:
Fig. 84
Fig. 85
Fig. 86
Fig. 87
Alarms assign the notification to an alarm list
or event list
Message text for the alarm / event list
Threshold of the alarm type
Default
No
selection
Inactive
Inactive
Inactive
Default
Inactive
Alarm
Channel X
threshold X
0,0000
Tabla de contenido
loading

Este manual también es adecuado para:

255201

Tabla de contenido