2.4.2 Battery Power

The scale can be operated on the internal rechargeable battery when AC power is not
available. The scale will automatically switch to battery operation if there is a power
failure or the power cord is removed.
Before using the scale for the first time, the battery should be fully
charged for up to 12 hours. During charging, the battery Indicator
displays the charge level (see Table 2-1), and the scale can be operated
during charging. The battery is protected against over charging and the
scale can remain connected to the AC power line.
Charge Status Indication Light
Off
Flashing Green
Flashing Yellow
Flashing Red
The scale can operate for approximately 60 hours on a fully charged battery.
During battery operation, the battery Indicator displays the battery status (see Table
2-2).
Charge Status Indication Light
Solid Green
Solid Yellow
Solid Red
Flashing Red
Flashing Red, Lo bAt
NOTE:
• Charging the scale must only be performed in a dry environment.
• Install the rubber seal securely to ensure the scale is protected when
operating in a wet environment.
CAUTION: Battery is to be replaced only by an authorized Ohaus service
dealer.
Risk of explosion can occur if replaced with the wrong type or connected
improperly.
Dispose of the lead acid battery according to local laws and regulations.
TABLE 2-1. BATTERY CHARGE STATUS.
TABLE 2-2. BATTERY CAPACITY STATUS.
Valor 2000 Series Scales
Charge level
Fully Charged
70 – 100 %
40 – 70 %
0 – 40 %
Capacity Remaining
70 – 100 %
40 – 70 %
10 – 40 %
0 – 10 %
Requires recharge;
Accuracy not guaranteed