Calibration Procedure; View The Calibration Data; Set The Calibration Reminder; Sample Measurements - Hach sensION+ MM340 Manual Del Usuario

Ocultar thumbs Ver también para sensION+ MM340:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
Option
Description
Data introduction
Manual probe constant introduction.
Theoretical calibration
The probe calibration data is replaced at 25 °C
(77 °F).

Calibration procedure

This procedure is for general use with liquid calibration solutions. Refer
to the documents that are included with each probe for additional
information.
Note: Solutions should be stirred during calibration. For more information about the
stirring settings, refer to
Change the stirring settings
1. Pour the buffers or calibration solutions into the labeled calibration
beakers.
2. From the main menu use the
CALIBRATION parameter. Confirm.
3. If required select the Operator ID (1 to 10) and confirm.
4. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
5. Push
to Start calibration.
6. Push
to measure the first calibration solution.
The next calibration solution is shown.
7. Rinse the probe with deionized water and put the probe into the
second calibration beaker. Be sure that there are no air bubbles in
the membrane.
8. Push
to measure the second calibration solution.
The next calibration solution is shown.
9. Rinse the probe with deionized water and put the probe into the third
calibration beaker. Be sure that there are no air bubbles in the
membrane.
10. Push
to measure the third calibration solution.
When the calibration is good, the display briefly shows Calibration
OK and then returns to the main menu.
Note: When a printer is connected the print menu opens and the result can be
printed.
10 English
on page 12.
or
and and to select the

View the calibration data

Data from the most recent calibration can be shown.
1. From the main menu use the
Confirm.
2. Select Display data.
3. Select Calibration data and confirm with
is shown.
• pH-the slope and offset values are shown alternating with the
deviation (in %) and calibration temperature.
• ORP-the measured mV value and calibration temperature are
shown.
• Conductivity-the cell constant and calibration temperature for each
standard are shown.

Set the calibration reminder

The calibration reminder can be set between 0 to 23 hours or 1-7 days
(default 1 day). The display shows the remaining time to the new
calibration.
Note: When 0 days is selected, the calibration reminder is turned off.
1. From the main menu use the
Confirm.
2. Use the
to enter the calibration menu.
3. Use the
or
to select Cal. frequency and confirm.
4. Use the and to advance to the next step and use the
change a value. Confirm.
Push
to start the calibration.

Sample measurements

Each probe has specific preparation steps and procedures for taking
sample measurements.
1. From the main menu use the
MEASURE. Confirm.
or
to select DATA LOGGER.
. The last calibration data
or
to select CALIBRATION.
or
or
and and to select
to
Tabla de contenido
loading

Tabla de contenido