Technical Specifications
Technical specifications
Measuring
frequency (Hz)
Main frequency
Voltage frequency
or parallel current
50,000-count Benchtop Multimeter
(cont.)
On this position, users can measure voltage fre quency.
Range
1 Hz to 5 Hz
50 Hz
500 Hz
5 kHz
50 kHz
500 kHz
1 MHz
Range
1 Hz to 5 kHz
5 kHz to 50 kHz
50 kHz to 500 kHz
500 kHz to 1 MHz
(∗) 500 mV available for 10 M
Measurement is performed by capacitive coupling.
Voltage range can be selected in manual mode by keeping the "Hz" key
pressed down.
Input resistance: 10 MΩ (11 MΩ on 5 V range)
Permanent admissible maximum voltage: 600 V rms
Varistance protection on voltage input.
Users can measure voltage frequency and magnitude.
Range
1 to 5 Hz
50 Hz
500 Hz
5 kHz
50 kHz
100 kHz
Range
1 Hz to 5 kHz
5 kHz to 50 kHz
50 kHz to 100 kHz
(∗) 1 kHz max. for 500 µA and 500 mV
(∗∗) 500 mV available for 10 M
Measurement is performed by capacitive coupling.
Input resistance: 10 MΩ (11 MΩ on 5 V range).
Permanent admissible maximum voltage: 600 Vrms
Accuracy
Resolution
0.03% + 2D
0.0001 Hz
0.03% + 2D
0.001 Hz
0.03% + 2D
0.01 Hz
0.03% + 2D
0.1 Hz
0.03% + 2D
1 Hz
0.05% + 2D
10 Hz
0.05% + 2D
100 Hz
Sensitivity (rectangular signals)
5 V (∗) to 500 V
2 % of rate
5 % of rate
10 % of rate
20 % of rate
Ω
inpur resistance
Accuracy
Resolution
0.03% + 5D
0.001 Hz
0.03% + 5D
0.01 Hz
0.03% + 5D
0.1 Hz
0.05% + 5D
1 Hz
0.05% + 5D
10 Hz
0.05% + 5D
100 Hz
Sensitivity (rectangular signals) Vrms
(∗∗) 5 V (∗) to 500 V
10% of rate
10% of rate
20% of rate
Ω
Ω
/ 1 G
input resistances
Admissible overload
1000 V
DC
or
700 V
AC
(1 min max.)
600 V
100 V
250 V
Admissible overload
1000 V
DC
or
700 V
AC
(1 min max.)
500 µA to 10 A
15 % of rate (∗)
15 % of max. rate
10 kHz
51