Functional Description (continued)
MATH
"MATH" mode
management
Ω
Func V, A,
, Hz
rAnGE
PArAM A
PArAM b
COEFF n, µ, m, K, M, G
Ω
Unit A to Z,
and Hz
MENU
"MENU" mode
management
bEEP On, bEEP OFF
SLEEP On, SLEEP OFF
Cont
bAud
SEtUP bASIC,
SEtUP uSER
rEJEC
Pt 100, Pt 1000
FILt On, FILt OFF
Ω
Ω
IMP 10M
, IMP 1G
46
" MATH " mode enables users to measure a given physical magnitude such as:
•
Volts
(0 - 10 V process, for example)
•
Amps
(4 - 20 mA current loop, for example)
•
Frequency (measuring rate or rotation speed, for example)
•
Ohms
(resistive position sensor, for example)
convert it and assign the adequate unit in order to obtain a direct reading of
the original magnitude on the instrument.
Keep the " MATH " key pressed down to access and exit this mode.
Six-phase programming is performed by successively pressing " MATH ",
then setting and validating the following via code wheel:
•
Definition of measured magnitude
•
Choice of measurement range
•
Definition of coefficient A of function y = Ax + b
•
Definition of coefficient B of function y = Ax + b (SI unit)
•
Choice of physical unit prefix to be displayed
•
Choice of physical unit to be displayed
The " MENU " mode enables users to define their own preferences for
certain instrument operation options and update internal clock date and
time.
Programming is performed in successive phases by pressing " MENU ",
then setting and validating the following via code wheel:
•
Validation or not of beep when pressing a key or during a warning while
unit is in operation.
•
Validation or not of automatic stop after 30 minutes operation if, during this
period, no action was performed on the front.
•
Step setting of LCD contrast (0, 33, 67, 100 %)
•
Parameterisation of transmission speed: 9600 / 19200 / 38400 / 57600.
Other transmission characteristics are fixed (8 data bits, 1stop bit, no
parity). The speed > 9,600 bauds is reserved preferably for data transfer.
•
When turning instrument on," bASIC " will allow you to find the
instrument's basic configuration, while " uSER " will show the last "
advanced " configuration set up by the user for each instrument mode;
this particularly allows you to adapt to " multi-user " instrument operation
(choose "bASIC" option).
•
50 Hz or 60 Hz: choice of network frequency used for optimal rejection.
Parameterisation of plate sensor type, 100 Ω ou 1000 Ω a 0°C, used to
•
measure temperature.
•
Select "FILt On" in order to measure the average value of alternating
signals via VDC mode.
•
Allows you to choose 500 mV range input impedance.
Functional Description
50,000-count Benchtop Multimeter