Saving, Retrieving And Deleting Results; Cleaning And Storing The Instrument; Rectifying Faults - Beurer BC 16 Manual De Instrucciones

Tensiómetro
Ocultar thumbs Ver también para BC 16:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 34
The WHO classification in the display shows the range of
the measured blood pressure.
If the values for systolic and diastolic pressure are in two dif-
ferent WHO ranges (e.g. systolic in the grade
sion range and diastolic pressure in the normal range), the
WHO classification on the unit always indicates the higher
range (grade
hypertension in the example described).

6. Saving, retrieving and deleting results

The computer automatically saves the blood pressure read-
ings of the last 50 measurements. When the 50 memory
locations are exceeded, the current oldest value is deleted.
You can display the memory as follows:
Press the User-Switching key
ory zone, and press the M key to switch from memory to
memory with sequence numbers.
The saved measured values are continuously stored.
Press the On/Off switch
If you forget to switch off the computer, it will switch off
automatically after 1 minute.
Press the User-Switching key
zone, then press and hold the M key for approximately
5-seconds to clear the data in the pre-designated memory
zone. The data in both memory zones – zone 1 and zone 2
may be erased if any of the batteries is removed.
hyperten-
to select a mem-
to turn the unit off.
to select a memory

7. Cleaning and storing the instrument

Clean your blood pressure monitor carefully using a
slightly damp cloth only.
Do not use any detergents or solvents.
Never hold the instrument under water as otherwise liquid
can penetrate and damage the instrument.
Never place any heavy objects on the instrument.

8. Rectifying faults

Error messages can occur when
the blood-pressure value is unusually high or low (
appears in the display),
you move or talk during the measurement (
the display),
the cuff tube is not properly inserted (
display),
inflation takes longer than 20 seconds (
display),
the inflation pressure is higher than 300 mmHg (
appears in the display),
an error occurs when storing the measured values (
appears in the display),
Exceeding measurement range (
In the about cases, you must repeat the measurement. Put
the batteries back in if necessary, or else replace them.
appears in
appears in the
appears in the
appears in the display).
Tabla de contenido
loading

Tabla de contenido