4 Acquiring CEPH Images
4.2.5
Locking bolt
Nasal positioner
Fixed bolt
Notch
CARPUS Plate
50
8. Click the READY button after the patient has been properly positioned. No
X-Ray will be emitted at this point.
9. Proceed to section 4.3 Initiating X-Ray Exposure.
CARPUS
Attaching the CARPUS plate
Positioning the Patient
1. Match the notch on the CARPUS plate
to the end of nasal positioner.
2. Slide the CARPUS plate onto the nasal
positioner.
3. Slide the locking bolt on the CARPUS
plate towards the nasal positioner and turn
it to secure the CARPUS plate in place.
4. Ensure that the CARPUS plate is firmly
in place.
1. Ask the patient to put his/her right hand
flat on the CARPUS plate. It is important to
ensure that the patient does not bend his/
her fingers.
2. Have the patient close their eyes and
remain still until scanning is complete.
PaX-i3D Green (PHT-60CFO) User Manual