Mid-sagittal plane
laser beam
Frankfurt plane laser
beam
Canine laser beam
Adjusting lever for the
canine laser
PaX-i3D Green (PHT-60CFO) User Manual
Aligning the Laser Beams
Be careful not to project the laser beams directly into the
patient's eyes as this could severely damage the patient's
WARNING
vision.
1. Mid-sagittal plane laser beam: Position the Mid-sagittal plane laser beam
at the center of the patient's face to prevent magnifications of the left or right
side in the final image.
2. Frankfurt plane laser beam: Position the patient's head so that the
Frankfurt plane is aligned with the Frankfurt plane laser beam.
3 Acquiring PANO Images
Frankfurt plane laser
beam UP button
Frankfurt plane laser beam
DOWN button
3. Canine laser beam: Have the patient
smile to properly position the canine laser
beam on the center of the patient's canine
tooth.