Troubleshooting - Gima 32924 Manual Del Usuario

Ocultar thumbs Ver también para 32924:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

TROUBLESHOOTING

SYMBOLS/SYMPTOMS
No power-on by pressing
button.
Measuring Error Symbol
appears when blood
pressure value displayed
is excessively low or high.
Measuring Error Symbol.
Measuring Error Symbol.
Measuring Error Symbol.
Note: If "EP" appears on the display, just return the device to your local distributor
or importer.
ENGLISH
CONDITIONS/CAUSES
Worn-out batteries.
Battery polarities have been
positioned incorrectly.
Cuff has been placed
incorrectly.
Did you talk or move during
measurement?
Shaking of the arm with
the cuff on.
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Inflation pressure exceeding
300 mmHg.
Error determining
measurement data.
INDICATION/CORRECTION
Replace them with 4 new
AA alkaline batteries.
Re-insert the batteries
in the correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again.
measurement.
Check cuff connection.
Measure again.
Switch the unit off,
then measure again.
Measure again.
36