Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below,
and should only be used in such environments:
IMMUNITY TEST
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
interruptions
and voltage variations
on power supply
input lines
IEC 61000-4-11
ENGLISH
IEC 60601
TEST LEVEL
3 A/m
supply lines
output lines
to line(s)
to earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
COMPLIANCE
LEVEL
3 A/m
supply lines
output lines
to line(s)
to earth
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
ELECTROMAGNETIC
ENVIRONMENT – GUIDANCE
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment.
Mains power quality should be
that of a typical commercial
or hospital environment.
If the user of the device requires
continued operation during
power mains interruptions,
it is recommended that
the device be powered from
an uninterruptible power supply
or a battery.